The application of detrended fluctuation analysis to assess physical characteristics of the human cell line ECV304 following toxic challenges

Open

Yasser Abd Djawad, David Attwood, Janice Kiely, Richard Luxton

2019 Sensing and Bio-Sensing Research Vol. 23 Article Cited by 5 Quartile

Abstract

In this paper we present a non-contact, impedance-based sensor system capable of characterizing the toxic response of cells to three different types of toxin. ECV304 cells were treated with 1 mM Hydrogen peroxide, 5% Dimethyl Sulfoxide, and 10 μg/ml saponin. Impedance spectroscopy was performed over a 2 h period on the cells within a commercial cell growth chamber, positioned on a pair of measurement electrodes, at frequencies between 200 and 830 kHz at 10 kHz intervals. Analysis of the impedance data was undertaken using the feature-extraction technique, Detrended Fluctuation Analysis (DFA). DFA scales the autocorrelation of a non-stationary signal, such as those generated using impedance spectroscopy for cytotoxicity testing. The correlation between the average fluctuation of the signal, F(n) (and scaling exponent, α) and a measurement of the cell size from image analysis was evaluated. The results showed that F(n) and α were strongly related to the changes of the morphological size of the cells. The results demonstrated that non-contact impedance spectroscopy, coupled with DFA can be used to monitor cell size in real time. © 2019

Affiliations

Electronics Department, Universitas Negeri Makassar, Indonesia; Institute of Biosensing Technology, University of the West of England, United Kingdom